FDADecember 16, 2020device

Philips EPIQ Diagnostic Ultrasound System; Model Numbers: EPIQ Elite, EPIQ 5G, EPIQ 5C, EPIQ 7G, EPIQ 7C, EPIQ CVx & EPIQ CVxi. MOD EPIQ Elite DIAGNOSTIC ULTRASOUND SYSTEM; Rx Only 100-240V~50/60 Hz, 600 VA MOD EPIQ 5 DIAGNOSTIC ULTRASOUND SYSTEM; GMDN 40761, Rx Only, 100-240V~50/60 Hz, 600VA MOD...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The manufacturer discovered an issue associated with the EPIQ Image Boost with xPlane and Color Flow or Doppler while using the X8-2t TEE Transducer. If Image Boost is enabled the color box shows flow visualization not accurately represent the fluid flow. Similarly, if Image Boost is enabled, both CW and PW Doppler traces will not accurately represent the fluid flow.

What to do

FDA enforcement status: Terminated

Brands named

philips ultrasoundphilips