FDAOctober 6, 2017device

Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6) (GTIN 00085412091570); Model Numbers: 35700BAX and 35700ABB. Intended to be used for the controlled administration of fluids.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Certain infusion pumps may exhibit excessive wear of the pumping mechanism caused by a lack of lubrication, this could affect the occlusion of the tubing and result in over- or under-infusion.

What to do

FDA enforcement status: Terminated

Brands named

baxter healthcarebaxter

UPCs

00085412091570

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.