Artis zee biplane MN, Model Number 10094143

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

What to do

FDA enforcement status: Ongoing

Brands named