FDAOctober 28, 2019device

Genius 2 Tympanic Thermometer OEM Tympanic Item Code: 303063

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (¿ 0.2¿C for Genius 2 and ¿ 0.3¿C for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermometers could exceed the upper stated accuracy tolerance of +0.2¿C for Genius 2 or +0.3¿C for Genius 3.

What to do

FDA enforcement status: Terminated

Brands named

cardinal health 200cardinalcardinal health