FDAFebruary 22, 2022device
ACL TOP 300 CTS/350 CTS/500 CTS/550 CTS/970 CL: Name Part Number 1.ACL TOP 300 CTS 00000280060 2. ACL TOP 300 CTS 00000280060R (Refurbished) 3. ACL TOP 350 CTS 00000280065 4. ACL TOP 500 CTS 00000280040 5. ACL TOP 500 CTS 00000280040R ( Refurbished) 6/ 6. ACL TOP 550 CTS 00000280045 7. ACL TOP 97...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and HemosIL Q.F.A. Thrombin (Bovine), specific to instrument models that utilize a single reagent probe
What to do
FDA enforcement status: Ongoing
Brands named
instrumentation laboratoryinstrumentation
UPCs
000002800600842695055691608426950928027000002800650842695078408100000280040084269504534990842695092800300000280045084269507292420000028009708430793045476
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAGEM Premier 5000; Part No. 00055415010.2026-01-20
- FDAGEM Premier 5000; Part No. 00055415008.2026-01-20
- FDAGEM Premier 5000; Part No. 00055415005.2026-01-20
- FDAGEM Premier 5000 PAK; Part No. 00055407504.2026-01-20
- FDAGEM Premier 5000 PAK; Part No. 00055407511.2026-01-20
- FDAGEM Premier 5000 PAK; Part No. 00055407508.2026-01-20
- FDAGEM Premier 5000 PAK; Part No. 00055407510.2026-01-20
- FDAGEM Premier 5000; Part No. 00055430004.2026-01-20