FDADecember 20, 2024device
AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx10mm UPN: M00553650
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
What to do
FDA enforcement status: Ongoing
Brands named
boston scientificboston
UPCs
08714729904595
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAOrca Single Use Air/Water and Suction Valves, Material Number (UPN) SUV-617-50; The Orca Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI (gastrointestinal) endoscopic procedure.2026-05-12
- FDAIntera Oncology INTERA 3000 Hepatic Artery Infusion Pump, AP-0300H, implantable drug delivery device2026-05-11
- FDABoston Scientific Model SC-4318, Clik X Anchor REF: M365SC431802026-04-29
- FDABoston Scientific CSK Electrodes: REF: CSK-TC10, CSK Electrode, STAINLESS STEEL, 10 cm x 28 ga (0.36 mm) REF: CSK-TC10-3M,CSK Electrode, STAINLESS STEEL, 3M Cable, 10 cm x 28 ga (0.36 mm); REF: CSK-TC15, CSK Electrode, STAINLESS STEEL, 15 cm x 28 ga (0.36 mm); REF: CSK-TC15-3M, CSK Electrode, STA...2026-04-29
- FDABoston Scientific TCN Electrodes: REF: TCN-5, TCN Electrode, NITINOL, 5 cm x 0.45 mm (0.0175 in); REF:TCN-10, TCN Electrode, NITINOL, 10 cm x 0.45 mm (0.0175 in) REF:TCN-15, TCN Electrode, NITINOL, 15 cm x 0.45 mm (0.0175 in); REF:TCN-20, TCN Electrode, NITINOL, 20 cm x 0.45 mm (0.0175 in) REF: T...2026-04-29
- FDALUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.52026-03-30
- FDALUX-Dx II Plus, Arrythmia detector and alarm, Model M312, with SERVER SW LATITUDE DRAGON EU (Europe), Model 6446, Version 7.5, or SERVER SW LATITUDE DRAGON AU (Australia/New Zealand), Model 6448, Version 7.52026-03-30
- FDABoston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P EL, Model Number U125; 2. VALITUDE CRT-P EL MRI, Model Number U128; 3. VISIONIST CRT-P EL, Model Number U225; 4. VISIONIST CRT-P EL, Model Number U226; 5. VISIONIST CRT-P EL MRI, Model Number U228.2026-03-19