FDADecember 21, 2016device

Fresenius 2008T Series Hemodialysis System

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine

What to do

FDA enforcement status: Terminated

Brands named

fresenius medical care renal therapiesfreseniusfresenius medical

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.