FDAFebruary 22, 2022device

Philips Hemodynamic Application, Model Number 722463, Software Version Number: R1.2.2 and R1.2.3. For visualization and analysis of patient hemodynamic data in the Philips Interventional Hemodynamic System.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

When a Neonatal patient is admitted, the Non-Invasive Blood Pressure is not set to Neonatal mode on the Philips Xperflex Cardio FC2010 patient monitor. If the NIBP is not set to Neonatal, the cuff inflation pressure applied will be higher than 150 mmHg and the deflation pressure will be higher than 5 mmHg. An applied cuff inflation pressure higher than 150mmHg or a deflation pressure higher than 5 could lead to bruising, mmHg nerve damage or bone deformity for a neonate.

What to do

FDA enforcement status: Terminated

Brands named

philips north americaphilipsphilips north

UPCs

00884838084544

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.