FDADecember 22, 2022device

MEDLINE UNITE FOOT & ANKLE Screw, Polyaxial Locking, Mini T8 (2.0x8mm), REF MMSL2008

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The recall is due to observed intra-operative screw failure. When pre-drilled with ¿1.3mm drill bit for ¿2.0mm screws and ¿1.6mm drill bit for ¿2.4mm we have observed screw failure due to excessive torque.

What to do

FDA enforcement status: Ongoing

Brands named

medline industries lp northfieldmedlinemedline industries

UPCs

10193489120707