FDAFebruary 6, 2019device

LW Scientific ZIP IQ PCV Centrifuge, ZIC-24HD-75T3 Rev C Product Usage: used as a general purpose centrifuge to spin down blood in hematocrit tubes

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The clear, plastic, polycarbonate hematocrit rotor may break apart during use and be ejected through the plastic housing, posing a safety hazard to the technician and potentially damaging the surrounding area.

What to do

FDA enforcement status: Terminated

Brands named

lw scientific

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →