FDASeptember 11, 2014device
MininephThl Human lgG4 Kit Product Code: ZK009.LR This kit is intended for qualifying human IgG subclasses 1,2,3 and 4 immunoglobulins in serum on the Beckman Coulter AU series. Meaurement of these immunoglobulins ais in the diagnosis of abnormal protein metabolism and the body's lack of ability ...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.
What to do
FDA enforcement status: Terminated
Brands named
binding sitebinding
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDABrand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXENT Analyser is an automated analyser intended for the qualitative and quantitative in vitro measurement of analytes in human body fluids used in conjunction w...2026-04-16
- FDAEmbrace Drill Tower, Wedged (25mm); Item Number: 645-081/62;2026-02-11
- FDAEmbrace Drill Tower, Standard/Lateral (25mm); Item Number: 645-081/63;2026-02-11
- FDAEndo-Model Replacement Plateau; Item Number: 15-0027/12;2026-01-12
- FDAEndo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/02;2026-01-12
- FDAEndo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/07;2026-01-12
- FDAEndo-Model Replacement Plateau; Item Number: 15-2835/12;2026-01-12
- FDAEndo-Model Replacement Plateau; Item Number: 15-2836/11;2026-01-12