FDADecember 4, 2017device

PORT-A-CATH II PS/Titan, Tray PAC II. Labeled as: a. VEN LP, PU, 2.6MM, UL, 8.5FR/INTRO (1/EA); b. DL, LP, PU, 2.2MM, WING-LOCK, 7FR/INTRO (1/EA); c. DL, LP, PU, 3.2MM, WING-LOCK, 10FR/INTRO (1/EA)

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Certain models and lots of PORT A CATH implantable port kits may have been misassembled. The kits may incorrectly contain a 6Fr or 6.5Fr size introducer instead of the correct 8.5Fr size introducer. The Power PORT A CATH II kit may also contain the incorrect locking collars in addition to the incorrect sized introducer.

What to do

FDA enforcement status: Terminated

Brands named

smiths medical asdsmithssmiths medical

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
PORT-A-CATH II PS/Titan, Tray PAC II. Labeled as: a. VEN LP, PU, 2.6MM, UL, 8.5FR/INTRO (1/EA); b. DL, LP, PU, 2.2MM, WING-LOCK, 7FR/INTRO (1/EA); c. DL, LP, PU, 3.2MM, WING-LOCK, 10FR/INTRO (1/EA) — Recall Details · AllClear