FDAJanuary 13, 2025device

Allura CV20 System Code: 722031

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel

What to do

FDA enforcement status: Ongoing

Brands named

philips medical systems nederland b vphilipsphilips medical