FDADecember 16, 2025device

Medline medical procedure convenience kits labeled as: 1) LAPAROSCOPY CDS-LF, REF CDS860235L; 2) GENERAL LAPAROSCOPIC, REF CDS930223M; 3) LAP CHOLE CDS, REF CDS983678K; 4) LAPAROSCOPY PELV CDS, REF CDS983679K; 5) LAP CHOLE CDS-LF, REF CDS983752M; 6) LAP CHOLE CDS-LF, REF CDS984016B; 7) LAP CHOLE ...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

What to do

FDA enforcement status: Ongoing

Brands named

medline industries lpmedlinemedline industries

UPCs

101984592586644019845925866510198459282607401984592826081019532757224240195327572243101953276415114019532764151210195327255855401953272558561019532719107840195327191079