FDAJanuary 5, 2017device

Spare Disk Remover Module; Product Number: 61001820 Intended for in vitro quantitative and qualitative determination of analytes in body fluids.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The firm became aware that the ball bushing axels are defective in the GSP instrument. During exposure to humid conditions, there is a risk that the axle material will corrode. Deterioration of the axle can be observed as deposits of corrosion on the defective axles. Ball bushing axles may corrode and corrosion particle may cause contamination in the wells or the corrosion will cause movement issues and possibly spillage and causing contamination of the plate wells.

What to do

FDA enforcement status: Terminated

Brands named

perkinelmer life and analytical sciences wallac oyperkinelmerperkinelmer life