FDAJanuary 28, 2019device

LivaNova Sorin Connect DPIII KOMPL. S5 REF 96-415-130, 2017-08-15, IPX1, RXOnly, (01)04033817901594 Product Usage: The S5 System is used to perform, control and monitor extracorporeal blood circulation during an operation for up to six hours.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Incorrect labeling; it was identified that some Datapad devices contained incorrect labels. Specifically the manufacturing date in the production identifier was incorrect and the manufacturing date on the unique device identifier label was labeled in the format of yy-mm-dd instead of yyyy-mm-dd.

What to do

FDA enforcement status: Terminated

Brands named

livanova

UPCs

04033817901594

Recall history