FDANovember 28, 2025device

SIGMA Spectrum Infusion Pump, Model/Catalog Number: 35700BAX2

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Certain pumps have potentially been released from service with defective grease applied to the cam and motor gears. Due to its low viscosity, the defective grease may lead to the device having insufficient lubrication on the cam, which could lead to premature wear of the mechanism assembly, resulting in over-infusion or a free-flow situation.

What to do

FDA enforcement status: Ongoing

Brands named

baxter healthcarebaxter

UPCs

00085412498683

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.