The LDL-Cholesterol assay is comprised of two distinct phases. In phase one a unique detergent solubilizes cholesterol from non-LDL- lipoprotein particles. This cholesterol is consumed by cholesterol esterase, cholesterol oxidase, peroxidase and 4- aminoantipyrine to generate a colorless end prod...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Beckman Coulter identified that LDL Cholesterol OSR6x96 is not meeting the performance claim listed in the Instructions for Use (IFU). LDL Cholesterol reagent showed clinically significant interference due to triglyceride concentration up to 1500 mg/dL may cause a maximum positive bias of up to 28.8% in low LDL Cholesterol samples (concentration 80 mg/dL) and up to 11.02% in high LDL Cholesterol samples (concentration 160mg/dL).
What to do
FDA enforcement status: Ongoing
Brands named
UPCs
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL LT SZ 5 CEM. Part Number: 150450105.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 5 CEM. Part Number: 150450205.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 4 CEM. Part Number: 150450204.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 7 CEM. Part Number: 150450207.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 3 CEM. Part Number: 150450203.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL LT SZ 7 CEM. Part Number: 150450107.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL LT SZ 4 CEM. Part Number: 150450104.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 8 CEM. Part Number: 150450208.2026-04-03
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →