FDAOctober 22, 2021device

ABX Pentra Calcium AS CP, Reagent 90 mL, REF A11A01954

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Product was shipped from a Medline warehouse to a customer at room temperature instead of refrigerated (ice). If affected product is used, the results from this impacted Calcium reagent will not be provided due to probable Quality Control failure. A delay in diagnosis may be experienced with repeat testing resulting in patient (customer) inconvenience, pain, and cost of the repeat blood draw and testing. All HII1220001954 product shipped/stored at a Medline branch between July 2021 and August 2021 in the identified lots were incorrectly stored and shipped. Six (6) boxes have been shipped to end users at room temperature instead of refrigerated (ice). Five (5) boxes are on hold at a Medline warehouse.

What to do

FDA enforcement status: Terminated

Brands named

medline industries lp northfieldmedlinemedline industries

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
ABX Pentra Calcium AS CP, Reagent 90 mL, REF A11A01954 — Recall Details · AllClear