FDAJanuary 18, 2021device

Tube Tracheostomy and Tube Cuff, Bivona¿ Cuffless Neonatal/Pediatric FlexTend" Tracheostomy Tube (model 60PFS45, lots 3991171, 3991172, and 3988103) V Neck Flange 4.5mm - Product Usage: intended to provide direct airway access for a tracheotomized patient for up to 29 days. It may be reprocessed ...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Smiths Medical became aware that three lot numbers of a specific model of Bivona¿ Cuffless FlexTend" Tracheostomy Tube (model 60PFS45, lots 3991171, 3991172, and 3988103) may have been incorrectly packaged with a straight neck flange instead of a V-neck flange. As a result, there is potential for patient injury if an incorrect device or accessory is selected for use. This use could lead to an internal injury resulting from excessive device contact with tissue, or an external patient injury due to external portions of the device pressing into or rubbing against tissue.

What to do

FDA enforcement status: Terminated

Brands named

smiths medical asdsmithssmiths medical

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.