FDAJuly 19, 2019device
Stryker, Omega 4.75 mm,. PEEK Knotless Anchor System, Double-Double, REF 3910-500-472, UDI 07613327464832, Sterile R, RX Only CE 0197
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Complaints have been received regarding difficulty in removing the driver from the screw. High driver removal forces can lead to the screw pulling out still attached to the driver. This could result in an additional anchor insertion site in the bone being used.
What to do
FDA enforcement status: Terminated
Brands named
stryker
UPCs
07613327464832
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDANeptune Smart Docking Station (120V), Catalog Number 0770-100-000, part of the Neptune Waste Management System2026-05-06
- FDADaig Livewire Steerable, Product Number 401582; REPROCESSED ELECTROPHYSIOLOGY CATHETER2026-04-10
- FDABARD Dynamic XT Steerable, Product Number 201104; REPROCESSED ELECTROPHYSIOLOGY CATHETER2026-04-10
- FDABARD Dynamic Deca Steerable, Product Number 201101; REPROCESSED ELECTROPHYSIOLOGY CATHETER2026-04-10
- FDABARD Dynamic Tip Steerable, Product Number 6DYNTP001;2026-04-10
- FDABARD Dynamic Tip Steerable, Product Number 200344; REPROCESSED ELECTROPHYSIOLOGY CATHETER2026-04-10
- FDABW Webster Duo-Decapolar, Product Number D728260RT; REPROCESSED ELECTROPHYSIOLOGY CATHETER2026-04-10
- FDADaig Livewire Steerable, Product Number 401603, REPROCESSED ELECTROPHYSIOLOGY CATHETER2026-04-10
Own something like this?