FDAJanuary 21, 2021device

OT1000 Series Orthopedic Surgical Tables

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

On January 7, 2021, STERIS identified that the glued bond between the beam and the leg spar assembly may begin to separate over time. This could result in unexpected movement of the affected leg spar if the separation occurred during a patient procedure.

What to do

FDA enforcement status: Terminated

Brands named

steris

Recall history