FDAFebruary 6, 2024device

Azurion 3 with a Certeray generator -To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722221 (2) 722222 (3) 722280

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.

What to do

FDA enforcement status: Ongoing

Brands named

philips medical systems nederland b vphilipsphilips medical