FDAFebruary 5, 2016device

Prime Series Stretchers (Fifth Wheel, Big Wheel and Zoom) Powered stretcher

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The hydraulic jack assemblies (PN: 0753-002-101, 0753-002-070, 0753-002-001, and 0753-002-001) were manufactured using the incorrect oil. The oil used does not contain the same wear resistance additives as the specified oils and therefore its use increases the potential for wear which may cause a leak event or jack drift. Testing was done to evaluate the jacks with incorrect oil and leak events we

What to do

FDA enforcement status: Terminated

Brands named

stryker medical divisionstrykerstryker medical

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Prime Series Stretchers (Fifth Wheel, Big Wheel and Zoom) Powered stretcher — Recall Details · AllClear