FDAJanuary 16, 2025device

Brand Name: RayStation Intended as a Radiation Therapy Treatment Planning System. Model/Catalog Number: 7.0.0.19, 8.0.0.61, 8.0.1.10 Software Version: RayStation 7, 8A and 8A Service Pack 1

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.

What to do

FDA enforcement status: Ongoing

Brands named

raysearch laboratories abraysearchraysearch laboratories