FDAFebruary 5, 2025device

IntelePACS (Image Fusion Module) - InteleViewer

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.

What to do

FDA enforcement status: Ongoing

Brands named

intelerad medical systemsinteleradintelerad medical