FDANovember 7, 2016device

Tosoh Bioscience HLC-723G8 Analyzer for HbA1c testing.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The device defect associated with this recall may result in HbA1c values with a negative bias in patients with the hemoglobin variants C, D, E and S. The negative bias occurs at the higher levels of HbA1c, e.g. 9%, but not at the diagnostic medical decision point of 6.5%.

What to do

FDA enforcement status: Terminated

Brands named

tosoh biosciencetosoh