FDANovember 2, 2025device
MiniMed 780G Insulin Pump Catalog Numbers: MMT-1884 MMT-1886 With Software Version 6.60 The MiniMed 780G insulin pump is indicated for use by patients age 7-80 years with Type 1 diabetes, whose total daily dose of insulin is 8 units per day or more. The MiniMed 780G system is intended for continu...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
A software timing and processor communication issue in MiniMed" 780G software version 6.60 may trigger Pump Error 43 and/or Pump Error 41, resulting in suspension of insulin delivery.
What to do
FDA enforcement status: Ongoing
Brands named
medtronic minimedmedtronic
UPCs
76300062988501
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAMedtronic O-arm O2 Imaging System. Model Number: BI70002000.2026-05-19
- FDAMedtronic 1x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64001; Stimulator, electrical, implanted, for parkinsonian tremor2026-05-13
- FDAMedtronic 2x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64002; Stimulator, electrical, implanted, for parkinsonian tremor2026-05-13
- FDAOctopus Evolution AS Tissue Stabilizer, Model TS25002026-05-12
- FDAOctopus Evolution Tissue Stabilizer, Model TS20002026-05-12
- FDAOctopus 4 Tissue Stabilizer, Model 294002026-05-12
- FDAMedtronic GUNDRY Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94115T; Cardiopulmonary bypass vascular catheter2026-04-21
- FDAMedtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94965; Cardiopulmonary bypass vascular catheter2026-04-21