Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6), product code 35700BAX - Product Usage: The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used for the controlled administration of fluids. These fluids may include pharmaceutical drugs, bloo...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
The flow rate test verification was not properly executed for certain Sigma Spectrum infusion pumps prior to release. The testing was performed for less than the required minimum run time. Improperly performed flow rate testing may not identify issues with pumps that could cause inaccuracies in flow rates which, in turn, could lead to over- or under- infusion. Potential risk to the patient depends on a number of factors including treatment and patient comorbidities. Depending on these factors, the patient could experience serious adverse health consequences.
What to do
FDA enforcement status: Terminated
Brands named
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAChlorthalidone Tablets, USP, 25 mg, [100 or 1000] Tablets pr bottle, Rx only, Manufactured by: Inventia Healthcare Limited, Additional Ambernath, M.I.D.C., Ambernath (East) - 421506, INDIA. Distributed by: Risiong Pharma Holdings, Inc., East Brunswick, NJ 08816. NDC 100-tablet bottle: 64980-599-0...2026-06-05
- FDASiemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay2026-05-07
- CPSCBayer Recalls 6 mL Size Afrin Original Nasal Spray Bottles Due to Risk of Serious Injury or Illness from Child Poisoning; Violates Mandatory Standard for Child-Resistant Packaging2026-04-30
- FDABrand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tubing sets with push-on connector on one end (to attach to the Fluid Selector Valve) and a spike or other appropriate connector on the opposite end to attach the source container.2026-04-22
- FDAepoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;2026-04-02
- FDABrand Name: Spacelabs Ultraview SL Command Module Product Name: SL Command Module Model/Catalog Number: 91496 Software Version: N/A Product Description: The Command Module (91496) is the core of the Spacelabs patient monitoring system, providing the processing power for all basic physiologic para...2026-03-26
- FDAGE Healthcare Omnipaque (iohexol) Injection, 300 mg/mL (64.7%), Rx Only, Distributed by: GE Healthcare Inc., Marlborough, MA 01752 USA, NDC 0407-1413-63.2026-03-24