FDAJanuary 29, 2024device

MRI system: Vantage Orian, Vantage Fortian, Model: MRT-1550

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.

What to do

FDA enforcement status: Ongoing