FDAFebruary 1, 2023device
Plato 17 Microcatheter: Straight Tip, PL17-160-000 ; 45 degree Tip, PL17-160-045; 90 degree Tip, PL17-160-090
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Microcatheters for embolic coil and diagnostic agent introduction to peripheral/neuro vasculature, due to manufacturing non-conformance, and due to flush, guidewire insertion and manipulation prior to placement, may have loose particles within catheter lumens that may be introduced to patients, which could cause embolic event, small vessel occlusion, and subacute to chronic inflammatory response.
What to do
FDA enforcement status: Ongoing
Brands named
scientia vascularscientia
UPCs
008180750104550081807501046200818075010479
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAArtegraft Vascular Graft; REF#: AG740;2026-02-10
- FDABroviac 4.2 Fr Single-Lumen CV Catheter Cutdown Tray with SureCuff Tissue Ingrowth Cuff Cat 600602025-12-12
- FDASafe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,2025-11-06
- FDAArtegraft¿ Collagen Vascular Graft; REF: AG1015;2025-09-25
- FDAArtegraft Collagen Vascular Grafts Model Numbers: (1) AG 616, (2) AG 630, (3) AG 636, (4) AG 640, (5) AG 645, (6) AG 715, (7) AG 730, (8) AG 735, (9) AG 740, (10) AG 745, (11) AG 750, (12) AG 830, (13) AG 840, (14) AG 1015, (15) AG 1030;2025-09-02
- FDAArtegraft Collagen Vascular Graft. Model/Catalog/Part Numbers: AG630M, AG636M, AG730M, AG740M, AG845M. The Artegraft is intended to serve as a substitute conduit for blood where bypass or replacement of occluded or diseased arterial segments is required or to establish a conduit for hemodialysis.2025-08-25
- FDAVenclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF Ablation Catheter prior to use. The Venclose digiRF Generator is a sophisticated multi-voltage energy delivery system featuring touchscreen control that automatically ...2025-08-21
- FDABrand Name: Q-Stop Product Name: Suture Retention Device Model/Catalog Number: QS Software Version: N/A Product Description: The DRS Vascular Q-Stop Suture Retention Appliance is a sterile, single-use, lightweight device made of biocompatible materials that can be easily installed within minutes ...2025-05-16