CDI Blood Parameter Monitoring System 500 Cardiopulmonary bypass on-line blood gas monitor Product Usage: The CDL System 500 provides continuous, on-line monitoring of the extracorporeal partial pressure of oxygen and carbon dioxide, pH, potassium, oxygen saturation, hematocrit, hemoglobin and te...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Specific CDI¿ Blood Parameter Monitoring System 500 devices are being voluntarily recalled because the BPM Sensor Head Assemblys Thermistor, which provides the blood temperature value that results in accurate display values on the monitor, does not meet specification. This may cause inaccurate temperature measurement and inaccurate analyte display values on the CDI System 500 monitor. Terumo Cardiovascular Systems (Terumo CVS) received complaints of inaccurate temperature measurements for specific devices distributed since November 5, 2015. These include both new production devices and recently serviced devices. POTENTIAL HAZARD There have been no reported illnesses or injuries as a result of this issue. A user who is not aware that the CDI System 500 is displaying inaccurate temperature values may not manage patient temperature appropriately. Using inaccurate information to manage warming and cooling strategies for a procedure could result in prolonged time on bypass. It could al
What to do
FDA enforcement status: Terminated
Brands named
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAPhilips Allura Xper with the following Model Numbers: Model # 722001 for Allura Xper FD10C; Model # 722002 for Allura Xper FD10F; Model # 722003 for Allura Xper FD10; Model # 722005 for Allura Xper FD10/10; Model # 722006 for Allura Xper FD20; Model # 722008 for Allura Xper FD20 Biplane; Model # ...2026-05-18
- FDAPhilips MultiDiagnost Eleva with the following Model Numbers: Model # 708032 for MultiDiagnost Eleva; Model # 708034 for MultiDiagnost Eleva with Flat Detector; Model # 708036 for MultiDiagnost Eleva; Model # 708037 for MultiDiagnost Eleva with Flat Detector; Model # 708038 for MultiDiagnost Elev...2026-05-18
- CPSCGenerac Power Systems Expands Recall of Generac and DR Power Electric Start Pressure Washers Due to Risk of Serious Injury or Death from Carbon Monoxide Hazard; Includes Additional Models2026-05-14
- FDANicotine Transdermal System Patch, 14 mg, 14 patches per box, Aveva, Distributed by: RUGBY LABORATORIES, Indianapolis, IN 46268, www.majorpharmaceuticals.com, NDC 0536-5895-88.2026-05-14
- FDAOctopus 4 Tissue Stabilizer, Model 294002026-05-12
- FDAOctopus Evolution Tissue Stabilizer, Model TS20002026-05-12
- FDAOctopus Evolution AS Tissue Stabilizer, Model TS25002026-05-12
- FDABrand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number: 4340 Software Version: N/A Product Description: The Inspire¿ Respiratory Sensing Lead (Model 4340) is designed to detect respiratory effort. The lead features a pressure sensitive membra...2026-05-11
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →