FDAMarch 20, 2019device

NM 830 Product Usage: The Discovery NM 630 is an all-purpose dual detector nuclear imaging system intended for general nuclear medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and ...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.

What to do

FDA enforcement status: Terminated