FDAFebruary 21, 2018device

F-T3SYSTEM, T3 (Trumpf Table Top) System The device is a surgical table accessory system used to allow prone positioning on OR table tops during spinal surgery. The system consists of a table top frame, arm supports,

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The recalling firm has become aware that the rail section clamps used to attach the leg support of the device may fail during use, which could result in critical harm to the patient.

What to do

FDA enforcement status: Terminated

Brands named

allen medical systemsallenallen medical

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
F-T3SYSTEM, T3 (Trumpf Table Top) System The device is a surgical table accessory system used to allow prone positioning on OR table tops during spinal surgery. The system consists of a table top frame, arm supports, — Recall Details · AllClear