Patient Side Cart, REF: 380620, with instrument arms, used as part of the da Vinci X Surgical System, REF: IS4200; Patient Side Cart, REF: 380652, with instrument arms, used as part of the da Vinci Xi Surgical System, REF: IS4000

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Preventative maintenance data was used to identify instrument Arms, part of surgical systems, that allow surgeons to perform surgical tasks, that may be at risk of insertion ballscrew failure, allowing for uncontrolled carriage movement along the insertion axis, which may cause tissue injury.

What to do

FDA enforcement status: Ongoing