FDAFebruary 5, 2016device

Vanguard XP Knee System Porous Femoral Component various sizes and orientation a total knee replacement system. prosthesis, knee, patella/femorotibial, semi-constrained, uncemented, porous, coated, polymer/metal/polymer

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The anterior/posterior (AP) dimension may be oversized by 0.016. If the femoral component does not fit as tightly as intended, bony ingrowth may not occur causing loosening of the femoral component. A revision surgery may be necessary.

What to do

FDA enforcement status: Terminated

Brands named

biomet

Recall history