FDAFebruary 17, 2015device

Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part number 03.010.048) and the Thumb Screw for the Aiming Arm for Titanium Cannulated Tibial Nails-EX (part number 03.010.052). The Recon Locking Aiming Arm for Lateral Entry Femoral Recon Nails- EX is used ...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

For certain lots, the incorrect raw material of annealed 17-4PH Stainless Steel was used instead of the correct material of heat-treated 304 Stainless Steel. This may lead to stress-corrosion cracking, which can lead to surgical delay or unintended debris entering the surgical site.

What to do

FDA enforcement status: Terminated

Brands named

synthes