FDAFebruary 16, 2016device

CIOS ALPHA (Image-intensified fluoroscope x-ray system) Product Usage: The Cios Alpha is a mobile X-ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluorosco...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Siemens found a defective part in their production line which could increase risk of an electrical shock for service personnel when working inside the device during system maintenance. The risk of an electrical shock is only present when a service technician ignores several steps listed in the service manual as precautionary actions during device maintenance.

What to do

FDA enforcement status: Terminated

Brands named

siemens medical solutionssiemenssiemens medical