FDAMarch 11, 2016device

Artis One, Interventional, Fluoroscopic, x-ray system Artis one is an angiography system developed for diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis One include cardiac angiography, neu...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

After importing segmentation results of the left atrium created on the Artis One system, the possibility exists for the results to be mirrored to an electro-anatomical 3D mapping system, CARTO from Biosense Webster Inc.

What to do

FDA enforcement status: Terminated