FDAOctober 16, 2013device

The HomeChoice Pro APD System is a peritoneal dialysis system. The HomeChoice Pro Automated Personal Cycler peritoneal dialysis system is intended for automatic control of dialysate solution exchange in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The occluder in affected devices 1) may have sharp edges that could cut the pneumatic tubing that connects the occluder blade to the manifold causing an unrecoverable system error, and/or 2) blade may separate from the back plate and cause Slow / No Flow alarm requiring therapy termination.

What to do

FDA enforcement status: Terminated

Brands named

baxter healthcarebaxter