FDADecember 15, 2017device

Uterine Electromyographic Monitor - Monica Novii Wireless Patch System Product Usage: The Novii Interface is an accessory to the Novii Pod which provides a means of interfacing the wireless output of the Novii Pod to the transducer inputs of a CTG Fetal monitor via three cables. The Novii Interfa...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

if the cables of the Monica Novii System are removed by a user, and then incorrectly re-connected the Fetal Heart Rate (FHR) and Maternal Heart Rate (MHR) signals would be displayed incorrectly (swapped).

What to do

FDA enforcement status: Terminated

Brands named

ge healthcare

Recall history