FDAMarch 22, 2019device

Cios Alpha with software version VA30-mobile C-arm x-ray Material # 11105200 Product Usage: The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to inte...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Continuous operation of the Cios Alpha at high tube outputs may result in increased wear of radiation emitting components, risk of premature failure of the Monoblock and loss of imaging X-rays exists

What to do

FDA enforcement status: Terminated

Brands named

siemens medical solutionssiemenssiemens medical

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.