system, x-ray, angiographic, model # 10094135, 10094137, 10094139, 10094141, 10280959 Artis zee/ zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

In Artis zeego systems, angulations in the vicinity of the C-arm collision area can cause the cable inlet to become clamped and can result in mechanical damage. In Artis systems with A100Plus generators, an attempt to resume operation following detection of a fault (such as a short circuit in the X-ray tube) can result in the failure of a module in the high-voltage generator.

What to do

FDA enforcement status: Terminated

Brands named

siemens medical solutionssiemenssiemens medical

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
system, x-ray, angiographic, model # 10094135, 10094137, 10094139, 10094141, 10280959 Artis zee/ zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. — Recall Details · AllClear