FDAFebruary 23, 2018device

GE Healthcare Centricity PACS-IW, Model Numbers: (a) 2052831-00X (b) 2049588-008 Product Usage: Centricity PACS-IW by GE Healthcare is a device that receives medical images (including mammograms) and data from various imaging sources. Images and data can be stored, communicated, processed and dis...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A database handling error could occur during the image acquisition process affecting the completeness of acquired images with Centricity PACS-IW. There is a potential that one or more image series (i.e. all images within an image set) may be missing from an exam without indication to the user. While this is rare, this can occur with imaging studies that consist of a very small number of images per series. E.g. CR Thorax exam with 1 image per series.

What to do

FDA enforcement status: Terminated

Brands named

ge healthcare