FDAMarch 3, 2015device

da Vinci Xi Surgical System, IS4000; ASSY,DISTAL SUJ,OUTER,IS4000, Material number: 380663-20 and 380663-21. The Intuitive Surgical Endoscope Instrument Control System (da Vinci Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endocopic Instrument...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Correction due to a fault that causes distal SUJ (DSUJ) failures that indicates a mismatch between the positions reported by the redundant position sensors on one of the patient cart arm joints.

What to do

FDA enforcement status: Terminated