FDAMarch 7, 2016device

VERO MHI-TM2000 Linear Accelerator System --- Common Name: Medical Linear Accelerator MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

In rare cases the VERO/MHI-TM2000 Operator Console could set an incorrect Gantry angle for the first beam of an IMRT treatment if certain specific conditions are met. If Gantry angle for the first beam of IMRT is incorrectly set and if nobody is aware of it, an erroneous treatment would be given to the patient.

What to do

FDA enforcement status: Terminated

Brands named

mitsubishi heavy industries ltd hiroshima machinery workmitsubishimitsubishi heavy

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
VERO MHI-TM2000 Linear Accelerator System --- Common Name: Medical Linear Accelerator MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated. — Recall Details · AllClear