FDANovember 17, 2017device

IntelliVue X3 Patient Monitor.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The NBP measurement of Intellivue X3 Patient Monitor shows intermittently only mean values instead of the diastolic and systolic blood pressure values. This is caused by a falsely detected NBP cuff, which is leading to wrong internal NBP setting. Furthermore, occasionally the Monitor shows the Check Touch Input message and the monitor is inoperable with the touch interface.

What to do

FDA enforcement status: Terminated

Brands named

philips electronics north americaphilipsphilips electronics