FDAMarch 7, 2018device

D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD 2.4MM X 10MM S-T LOCKING SCREW, REF 74692410, QTY (1), STERILE R

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch.

What to do

FDA enforcement status: Terminated

Brands named

smith nephewsmithsmith nephew