FDAFebruary 13, 2025device

Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5S-01C, (2) G5S-02C, (3) G5S-10C, (4) G5S-11C; Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such a...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

What to do

FDA enforcement status: Ongoing

Brands named

zoll medicalzoll

UPCs

00812394021246

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5S-01C, (2) G5S-02C, (3) G5S-10C, (4) G5S-11C; Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such a... — Recall Details · AllClear