FDAJanuary 30, 2024device
MINI STICK MAX 4F X 10 CM STD .018 SS/PD ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-749
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
What to do
FDA enforcement status: Ongoing
Brands named
angiodynamics
UPCs
15051684022927
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDASoft-Vu Angiographic Catheter, Sos Omni Selective (1), Non-Braided; SOFT-VU SO1 4F X 80CM 038 NB 0SH; Catalog No.: 10714024; Product/UPN No.: H787107140245 (Box), H787107140240 (Pouch); Box Quantity: 5 pouches;2026-03-03
- FDASoft-Vu Angiographic Catheter, Cobra (1), Non-Braided; SOFT-VU CB1 4F X 100CM 038 NB 0SH; Catalog No.: 10714017; Product/UPN No.: H787107140175 (Box), H787107140170 (Pouch); Box Quantity: 5 pouches;2026-03-03
- FDASoft-Vu Angiographic Catheter, Cobra (2), Non-Braided; SOFT-VU CB2 4F X 100CM 038 NB 0SH; Catalog No.: 10714018; Product/UPN No.: H787107140185 (Box), H787107140180 (Pouch); Box Quantity: 5 pouches;2026-03-03
- FDASoft-Vu Angiographic Catheter, Pigtail, Non-Braided; SOFT-VU PT 4F X 100CM 035 NB 10SH; Catalog No.: 10707103; Product/UPN No.: H787107071035 (Box), H787107071030 (Pouch); Box Quantity: 10;2026-03-03
- FDASoft-Vu Angiographic Catheter, Pigtail, Non-Braided; SOFT-VU PT 4F X 65CM 038 NB 10SH; Catalog No.: 10714002; Product/UPN No.: H787107140025 (Box), H787107140020 (Pouch); Box Quantity: 10 pouches;2026-03-03
- FDASoft-Vu Angiographic Catheter, Kumpe, Non-Braided; SOFT-VU KMP 4F X 65CM 038 NB 0SH; Catalog No.: 10714014; Product/UPN No.: H787107140145 (Box), H787107140140 (Pouch); Box Quantity: 5 pouches;2026-03-03
- FDASoft-Vu Angiographic Catheter, Omni Flush, Non-Braided; SOFT-VU OF 4F X 65CM 038 NB 6SH; Catalog No.: 10714001; Product/UPN No.: H787107140015 (Box), H787107140010 (Pouch); Box Quantity: 10 pouches;2026-03-03
- FDASoft-Vu Angiographic Catheter, Headhunter (1), Non-Braided; SOFT-VU HH1 4F X 100CM 038 NB 0SH; Catalog No.: 10714040; Product/UPN No.: H787107140405 (Box), H787107140400 (Pouch); Box Quantity: 5 pouches;2026-03-03